Read me file for HI-PRO Update Ver. 1.11 for Windows.
-----------------------------------------------------------------------------

Date : 12'th July 1999                                 Madsen Electronics

History of changes :
=============================================================================
Ver. 1.00  first version.

Ver. 1.01  

- Potential error which might crash the PC corrected.
- Screen flicker removed at startup.

New debugging options :

-  	When running the Update program all activity including error messages are
	now logged to a file "WUpdate.log" in the Windows directory.
-  	The log file can be displayed if selecting the "View logfile" item in
	the system menu.

-  	The HIPRO ini file can be viewed/changed by selecting the "View HIPRO.INI"
	item the system menu.

Ver.1.02 

-  	If the current directory was changed while running HI-PRO update,
	it could not find its hex files. This is now fixed, the hex files
	are now always read from the directory in which the HI-PRO exe file is.

Ver.1.03 

-	A downgrade warning is issued when downloading a older version to
	HI-PRO than present.
-	When starting up, the HI-PRO main window is displayed instantly, before
	it was only shown when a HI-PRO was detected.

Ver. 1.04 

-	Multilanguage support. Uses the HIPRO.INI language setup.
-	Improved error handling. HI-PRO will prompt the user to start
        the HI-PRO test program if the HI-PRO could not be found.

Ver. 1.05 

-	Improved error handling when defective hex files.
-	Downgrade now not possible unless the parameter "/D" 
        is issued when invoking the program.

Ver. 1.06 

-	Fixed problem downloading a new firmware, if no valid firmware
	was already in HI-PRO.

Ver. 1.07

-	Fixed problem downloading a new firmware, if HI-PRO reports 
	selftest error.

Ver. 1.08

-	Fixed "General protection error" on some computers when downloading 
	a new firmware (Stack increased).

Ver. 1.09

-	Now uses the HI-PRO driver version 2.20, updated the support
        dll's and HI-PRO test program. 

Ver. 1.10

-	Now uses the HI-PRO driver version 2.23, updated the support
        dll's and HI-PRO test program. 

Ver. 1.11

-	Now uses the HI-PRO driver version 2.25, updated the support
        dll's and HI-PRO test program. 



History log end.
===============================================================================
The HI-PRO Update program (WUpdate.exe) is used to update the firmware in the
HI-PRO unit.

The HI-PRO DLL driver files must be installed in the directory
where the Update program resides. A valid HIPRO.INI file
must reside in the Windows directory.

Firmware version supplied  :

  2.20 (HIPRO220.HEX)

When the Update program starts up, it reads the header in all the HEX files
in current directory to identify the version. The version found is then added to
a selection list if the monitor version in the file matches with the monitor
version in HI-PRO.
If you want to add a new HI-PRO version to the selection list, you should
copy the new HEX file to the current directory.

The Update program is able to run in three different startup modes :

1. Manual mode.
	The Update program is launched with no command line parameters.
	The user is able to choose firmware version.
	The user must press the "Start update" button to start the update
	process.

2. Semi manual mode.

	The Update program is launched with the "/FFilename.HEX" command
	line parameter, "Filename.HEX" is a file containing the wanted HI-PRO
	version.
	The user will NOT be able to choose firmware version.
	The user must press the "Start update" button to start the update
	process.
	Example : WUpdate.exe /FHIPR109.HEX

3. Auto mode.

	The Update program is launched with the "/A" command line parameter.
	Notice, the "/A" parameter is intended to be used together with
	the "/FFilename.HEX" parameter.
	Updating is started automatically, when it is finished, a status
	message is displayed and the user will only be able to exit the 
        program.
	Example : WUpdate.exe /FHIPR108.HEX /A

Downgrading of HI-PRO:

If choosing a older firmware version than present in HI-PRO, the 
"Start update" button will be grayed out (disabled). 
This protection can be bypassed by using the "/D" parameter when
invoking the program.

Example : WUpdate.exe /D

------------------------------------------------------------------------
Hints for programmers :

If the Update program is launched from a fitting program, be sure to
to logout from HI-PRO (if you are logged in) before launching, and do
not login until the update program has terminated.